Interlaboratory concordance in HER2 testing: Results of a Serbian ring-study
Authors/Creators
- 1. Oncology Institute of Vojvodina, Sremska Kamenica, Serbia
- 2. 2Institute of Oncology and Radiology of Serbia, Belgrade, Serbia;
- 3. 2Institute of Oncology and Radiology of Serbia, Belgrade, Serbia
- 4. Medical Faculty, University of Nis, Clinical Center Nis, Serbia
- 5. Institute of Pathology University of Belgrade, Faculty of Medicine, Belgrade, Serbia
- 6. Clinical Center Kragujevac, Kragujevac, Serbia
- 7. Medical Center "Bezanijska Kosa", Belgrade, Serbia.
Description
urpose: The purpose of this study was to assess the immunohistochemistry and chromogenic in situ hybridization (CISH) interlaboratory consensus between national pathology laboratories in Serbia.Methods: This study was conducted between 2013 and 2016. In 2013, HER2 results were evaluated using two sets of four different breast cancer specimens in five laboratories. A total of 20 immunohistochemistry and 20 CISH cases were tested. In 2014, there were 6 testing rounds, and a total of 24 specimens were analyzed, whereas in 2015 and 2016, seven testing rounds were conducted, with four additional cases (i.e. a total of 28 specimens). In 2014, 2015 and 2016, all institutions performed immunohistochemical analysis only.Results: We found discrepancies in HER2 immunohistochemical (IHC) results in all four surveys. IHC testing resulted in diagnostic discordance between participating centers in two (2/17) cases in 2013, two (2/24) in 2014, four (4/27) cases in 2015 and three cases (3/27) in 2016. The overall agreement among the centers was 79%, 85.5%, 83.5% and 89.4%, respectively. For CISH analyses, the results for 16 (84.2%) of 19 samples were consistent for all participants. Three results were found to be discordant, indicating a misdiagnosis rate of 15.8%. In all the discrepant cases, interinstitutional discordances were related to technical and evaluation issues.Conclusions: Our study highlights the difficulty encountered during HER2 testing using immunohistochemistry and CISH. This also emphasizes the need for rigorous quality control procedures for specimen preparation and analysis.
Files
Interlaboratory concordance in HER2 testing. Results of a Serbian ring study, 24-3-1045.pdf
Files
(948.6 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:93898b49ef7075bdfe7a33fb0161ef0c
|
948.6 kB | Preview Download |
Additional details
Related works
- Is part of
- 1107-0625 (ISSN)
- 2241-6293 (ISSN)