Calibration of Syringe Pumps Using Interferometry and Optical Methods

Description

Syringe pumps are commonly used for drug delivery
in hospitals and clinical environments. These instruments are critical
in neonatology and oncology, where any variation in the flow rate
and drug dosing quantity can lead to severe incidents and even death
of the patient. Therefore it is very important to determine the
accuracy and precision of these devices using the suitable calibration
methods. The Volume Laboratory of the Portuguese Institute for
Quality (LVC/IPQ) uses two different methods to calibrate syringe
pumps from 16 nL/min up to 20 mL/min. The Interferometric method
uses an interferometer to monitor the distance travelled by a pusher
block of the syringe pump in order to determine the flow rate.
Therefore, knowing the internal diameter of the syringe with very
high precision, the travelled distance, and the time needed for that
travelled distance, it was possible to calculate the flow rate of the
fluid inside the syringe and its uncertainty. As an alternative to the
gravimetric and the interferometric method, a methodology based on
the application of optical technology was also developed to measure
flow rates. Mainly this method relies on measuring the increase of
volume of a drop over time. The objective of this work is to compare
the results of the calibration of two syringe pumps using the different
methodologies described above. The obtained results were consistent
for the three methods used. The uncertainties values were very
similar for all the three methods, being higher for the optical drop
method due to setup limitations.