Evaluation of Sphygmomanometer Dial Performance Across Variable Temperatures and Pressure Conditions
Authors/Creators
- 1. Research Center for Equipment Manufacturing Technology, Management Building 720, B.J. Habibie Science and Technology Area, South Tangerang, Banten, 15314, National Research and Innovation Agency, Republic of Indonesia
- 2. Research Center for Pharmaceutical Ingredients and Traditional Medicine, Meatpro Building, Soekarno Science and Technology Area, Jakarta - Bogor Highway KM 46, Cibinong West Java 16911, National Research and Innovation Agency, Republic of Indonesia
- 3. Research Center for Electrical Technology, Management Building 720, B.J. Habibie Science and Technology Area, South Tangerang, Banten, 15314, National Research and Innovation Agency, Republic of Indonesia
Contributors
Contact person:
- 1. Research Center for Equipment Manufacturing Technology, Management Building 720, B.J. Habibie Science and Technology Area, South Tangerang, Banten, 15314, National Research and Innovation Agency, Republic of Indonesia
Description
In this study, we evaluated the performance of aneroid sphygmomanometers dials under specific temperatures. Five different Sphygmomanometer dials were placed inside the chamber. The dials were kept in the chamber with five different temperature settings for 1 hour, and the pressures were applied to the dial from 0 to 300 mmHg and vice versa. The result shows that the performance of all the dial models based on a one-factor analysis of variance has shown that the parameters have significant effect. The first variable affecting the performance is the pressure and model. Based on the GR&R using the ANOVA method, the total GR&R based on % Combination Variance is 67.87% and % Study Variance is 82.38. The most significant sphygmomanometer dial model that affected the treatment is model C. This study indicates that based on IEC 80801 or by the GR&R, there are potential measurement consistency and accuracy issues, so the manufacturer requires some investigation and corrective action.
Notes
Files
p364-372.pdf
Files
(1.3 MB)
| Name | Size | Download all |
|---|---|---|
|
md5:c70e7281c2170018624b4a34c6a00b9a
|
1.3 MB | Preview Download |
Additional details
Related works
- Is identical to
- Journal article: 10.5109/7411071 (DOI)
- Is supplemented by
- Other: https://citation.crossref.org/?doi=10.5109/7411071 (URL)