Journal article Open Access

DEVELOPMENT AND VALIDATION OF A STABILITY INDICATING RP-HPLC METHOD FOR THE DETERMINATION OF TERIZIDONE

Santosh V. Gandhi*, Vaishali P. Shevale and Ganesh B. Choudhari

A simple, rapid, sensitive, and linear stability indicating reverse phase HPLC method was developed for estimation of Terizidone. In this method separation was carried out on HiQSil C8 column (250mm*4.6 mm, 5μm) using Ammonium Acetate Buffer pH 3 (adjusted with glacial acetic acid) and methanol (60:40 v/v) as mobile phase at flow rate of 1 ml/min. The quantification was carried out at 264 nm. The retention time (tR) of drug was 7.3± 0.10 min. The method was validated with respect to linearity, precision, assay, accuracy and robustness. The data of linear regression analysis indicated a good linear relationship over the range of 5-30 μg/ml concentrations with a correlation coefficient (r 2 ) of 0.9837. Terizidone was subjected to different stress testing conditions. The degradation products were well resolved from the drug under the tested conditions. Keywords: Terizidone, High performance liquid chromatography, stability indicating method, validation.

Files (1.0 MB)
Name Size
6.Manuscript_IAJPS.pdf
md5:0ccdb3808286ed8d7ea878bd43dbccca
1.0 MB Download
7
3
views
downloads
All versions This version
Views 77
Downloads 33
Data volume 3.1 MB3.1 MB
Unique views 77
Unique downloads 33

Share