In vivo Histomorphometric and Corrosion Analysis of Ti-Ni-Cr Shape Memory Alloys in Rabbits
Description
A series of Ti based shape memory alloys with
composition of Ti50Ni49Cr1, Ti50Ni47Cr3 and Ti50Ni45Cr5 were
developed by vacuum arc-melting under a purified argon atmosphere.
The histometric and corrosion evaluation of Ti-Ni-Cr shape memory
alloys have been considered in this research work. The alloys were
developed by vacuum arc melting and implanted subcutaneously in
rabbits for 4, 8 and 12 weeks. Metallic implants were embedded in
order to determine the outcome of implantation on histometric and
corrosion evaluation of Ti-Ni-Cr metallic strips. Encapsulating
membrane formation around the alloys was minimal in the case of all
materials. After histomorphometric analyses it was possible to
demonstrate that there were no statistically significant differences
between the materials. Corrosion rate was also determined in this
study which is within acceptable range. The results showed the Ti-
Ni-Cr alloy was neither cytotoxic, nor have any systemic reaction on
living system in any of the test performed. Implantation shows good
compatibility and a potential of being used directly in vivo system.
Files
16418.pdf
Additional details
References
- <p>
- S.K. Wu, H.C. Lin, T.Y. Lin Electrical resistivity of Ti–Ni binary and Ti–Ni–X (X = Fe, Cu) ternary shape memory alloys, Materials Science and Engineering A 438–440 (2006) 536–539
- HuibinXu, Chengbao Jiang, Shengkai Gong, Gen Feng, Martensitic transformation of the Ti50Ni48Fe2 alloy deformed at different temperatures, Materials Science and Engineering A.281 (2000) 234–238
- C. Li, Y.F. Zheng ,The electrochemical behavior of a Ti50Ni47Fe3 shape memory alloy, Materials Letters 60 (2006) 1646–1650
- Tan Yong-heng, U Hong-hui, DU Yong, Isothermal section at 927C of Cr-Ni-Ti system, Trans. Nonferrous Met. SOC. China 17(2007) 71 1- 714
- SiaNemat-Nasser, Jeom Yong Choi, Strain rate dependence of deformation mechanismsin a Ni–Ti–Cr shape-memory alloy, ActaMaterialia 53 (2005) 449–454
- J. Uchil, K. Ganesh Kumara, K.K. Mahesh, Effects of heat treatment temperature and thermal cycling on phase transformations in Ni–Ti–Cr alloy, Journal of Alloys and Compounds 325 (2001) 210–214
- ASTM F86 – 12: Standard Practice for Surface Preparation and Marking of Metallic Surgical Implants, Annual Book of ASTM Standards. Philadelphia, Pennsylvania, USA: American Society for Testing and Materials; 2004.
- Bancroft J.D. and M. Gamble, Theory and Practice of Histological Techniques. 5th Ed; Churchill Livingstone, London, UK, (2007) 125- 138.
- F. Witte, J. Fischer, J. Nellesen, C. Vogt, J. Vogt, T. Donath, F. BeckmannIn vivo corrosion and corrosion protection of magnesium alloy LAE442, ActaBiomaterialia 6 (2010) 1792–1799 [10] JormaRyhänen,Biocompatibilityevaluation of Nickel Titaniumshape memorymetal alloy, (PhD Thesis) Oulu University Library,Oulu 1999, 10.</p>