Noise Performance Optimization of a Fast Wavelength Calibration Algorithm for OSAs
Authors/Creators
Description
A new fast correlation algorithm for calibrating the
wavelength of Optical Spectrum Analyzers (OSAs) was introduced
in [1]. The minima of acetylene gas spectra were measured and
correlated with saved theoretical data [2]. So it is possible to find the
correct wavelength calibration data using a noisy reference spectrum.
First tests showed good algorithmic performance for gas line spectra
with high noise. In this article extensive performance tests were made
to validate the noise resistance of this algorithm. The filter and
correlation parameters of the algorithm were optimized for improved
noise performance. With these parameters the performance of this
wavelength calibration was simulated to predict the resulting
wavelength error in real OSA systems. Long term simulations were
made to evaluate the performance of the algorithm over the lifetime
of a real OSA.
Files
14021.pdf
Additional details
References
- Fuhrmann, T. "Fast Wavelength Calibration Algorithm for Optical Spectrum Analyzers". International Journal of Information and Communication Engineering 6:1 2010, pp. 1-6.
- S. L. Gilbert, W. C. Swann "Acetylene 12C2H2 Absorption Reference for 1510 nm to 1540 nm Wavelength Calibration - SRM 2517a". NIST Special Publication 260-133, 2001.
- ITU-T "Spectral Grids for WDM applications: DWDM frequency grid". ITU-T, G694.1, 06/2002.
- K. Benzel, A. Conzelmann, T. Fuhrmann, E. Löcklin "Device for simultaneous analysis of multiple optical channels". European Patent EP 1106979, 1999.
- JDSU "ONT Platforms, Optical Spectrum Analyzer Modules". JDSU Product Overview.
- A. Savitzky, M. J. E. Golay "Smoothing and Differentiation of Data by Simplified Least Square Procedure". Analytical Chemistry, Vol. 36, No. 8, pp. 1627-1639, 1964.