In this paper, algorithm estimating the blood pressure
was proposed using the pulse transit time (PTT) as a more convenient
method of measuring the blood pressure. After measuring ECG and
pressure pulse, and photoplethysmography, the PTT was calculated
from the acquired signals. Thereafter, the system to indirectly measure
the systolic pressure and the diastolic pressure was composed using
the statistic method. In comparison between the blood pressure
indirectly measured by proposed algorithm estimating the blood
pressure and real blood pressure measured by conventional
sphygmomanometer, the systolic pressure indicates the mean error of
±3.24mmHg and the standard deviation of 2.53mmHg, while the
diastolic pressure indicates the satisfactory result, that is, the mean
error of ±1.80mmHg and the standard deviation of 1.39mmHg. These
results are satisfied with the regulation of ANSI/AAMI for
certification of sphygmomanometer that real measurement error value
should be within the mean error of ±5mmHg and the standard
deviation of 8mmHg. These results are suggest the possibility of
applying to portable and long time blood pressure monitoring system
hereafter.
Files
11733.pdf
Additional details
References
- K. G. Ng and C. F. Small, "Survey of automated noninvasive blood pressure monitors", J. Clin. Eng., Vol. 19, pp.452-487, 1994.
- R. P. Kelly, C. S. Hayward, A. P. Avolio, M. F. O'Rourke, "Noninvasive determination of age-related changes in the human arterial pulse", Circulation, Vol. 80, pp.1652-1659, 1989.
- E.K. Park, S.M. Lee, Y.H. Han, et al., "A study on estimation of systolic blood pressure using PTT", J. Biomed. Eng. Res., Vol. 25, No. 6, pp. 605-609, 2004.
- J. Lass, K. Meigas, D. Karai, et al., "Continuous blood pressure monitoring during exercise using pulse wave transit time measurement", Proceedings of the 26th Annual International Conference of the IEEE EMBS, pp.2239-2242, 2004.
- X. F. Teng and Y. T. Zhang, "Continuous and Noninvasive Estimation of Arterial Blood Pressure Using a Plethysmographic Approach", Proceedings of the 25th Annual International Conference of the IEEE EMBS, pp.3153-3156, 2003.
- G. V. Marie, C. R. Lo and D. W. Johnston, "The relationship between pulse transit time and blood pressure", Psycholphysiology, Vol. 21, No. 5, pp.521-7, 1984.
- Gi Ryon Kim, Gwang Nyeon Lim, Byeong Cheol Choi, et al., "Implementation and evaluation of the sensor assessing pressure and photoplethysmogram", J. of the Korean Sensors Society, Vol. 15, No. 2, pp.00-00, 2006.