Published December 25, 2009 | Version 1653

Development of a Non-invasive System to Measure the Thickness of the Subcutaneous Adipose Tissue Layer for Human

Description

To measure the thickness of the subcutaneous adipose tissue layer, a non-invasive optical measurement system (λ=1300 nm) is introduced. Animal and human subjects are used for the experiments. The results of human subjects are compared with the data of ultrasound device measurements, and a high correlation (r=0.94 for n=11) is observed. There are two modes in the corresponding signals measured by the optical system, which can be explained by two-layered and three-layered tissue models. If the target tissue is thinner than the critical thickness, detected data using diffuse reflectance method follow the three-layered tissue model, so the data increase as the thickness increases. On the other hand, if the target tissue is thicker than the critical thickness, the data follow the two-layered tissue model, so they decrease as the thickness increases.

Files

1653.pdf

Files (482.8 kB)

Name Size
md5:7cd43ce6c92337d835ed5dae7eff0f3d
482.8 kB Preview Download

Additional details

References

  • Durnin JVGA, Rahaman MM. The assessment of the amount of fat in the human body from measurements of skin fold thickness. Br J Nutr 1967;21:681-9
  • Heymsfield SB, McMannus C, Smith J, Stevens V, Nixon DW. Anthropometric measurement of muscle mass: revised equations for calculating bone-free arm muscle area. Am J Clin Nutr 1982;36:680-90
  • Borkan GA, Gerzof SG, Robbins AH, Hults DE, Silbert CK, Silbert JE. Assessment of abdominal fat content by computed tomography. Am J Clin Nutr 1982;36:172-177
  • Borkan GA, Hults DE, Gerzof SG, Robbins AH. Comparison of body composition in middle-aged and elderly males using computed tomography. Am J Phys Anthropol 1985;66:289-95
  • Lukaski HC. Methods for the assessment of human body composition: traditional and new. Am J Clin Nutr 1987;46:537-56
  • Booth RAD, Goddard BA, Paton A. Measurement of fat thickness in man: a comparison of ultrasound, Harpenden calipers and electrical conductivity. Br J Nutr 1966;20:719-725
  • Jones PRM, Davis PSW, Norgan NG. Ultrasonic measurements of subcutaneous adipose tissue thickness in man. Am J Phys Anthropol 1986;71:359-363
  • Möller R, Tafeit E, Smolle KH, Kulling P. "Lipometer": determining the thickness of a subcutaneous fatty layer. Biosens Bioelectron 1994;9:xiii-xvi
  • Brozek J, Kinzey W. Age changes in skinfold compressibility. J Gerontol 1960;15:45-51 [10] Himes JH, Roche AF, Siervogel RM. Compressibility of skinfolds and the measurement of subcutaneous fatness. Am J Clin Nutr 1979;32:1734-1740 [11] Brozek J. Body measurements, including skinfold thickness, as indicators of body composition. In: Brozek J, Henshel A, eds. Techniques for measuring body composition. Washington, DC: National Academy of Sciences-National Research Council, 1961: 3-35 [12] Markman B, Barton FE Jr. Anatomy of the subcutaneous tissue of the trunk and lower extremity. Plast Reconstr Surg 1987;80(2):248-54. [13] Desch├¬nes D, Couture P, Dupont P, Tchernof A. Subdivision of the subcutaneous adipose tissue compartment and lipid-lipoprotein levels in women. Obes Res 2003;11(3):469-476