Orbitrap Exploris 480 Method Summary Global Settings Method Duration (min)= 11 Use Ion Source Settings from Tune = Not Checked Use Static Source Gasses Collision Gas Type = -1 Collision Gas pressure (mTorr)= 1 Ion Source Type = H-ESI Spray Voltage = Static Spray Voltage: Positive Ion (V) = 3500 Spray Voltage: Negative Ion (V) = 2500 Sheath Gas (Arb) = 60 Gas Mode = Static Infusion Mode (LC)= False Aux Gas (Arb) = 15 Sweep Gas (Arb) = 2 Ion Transfer Tube Temp (°C) = 350 Vaporizer Temp (°C) = 400 APPI Lamp = Not in use FAIMS Mode = Not Installed Application Mode = Small Molecule Pressure Mode = Standard Expected Peak Width (s) = 4 Default Charge State = 1 Advanced Peak Determination = True Xcalibur AcquireX enabled for method modifications = True Mild Trapping = False Experiment 1 Start Time (min) = 0 Experiment Name = MS End Time (min) = 11 Scan MasterScan Orbitrap Resolution = 15000 Scan Range (m/z) = 100-1000 Microscans = 1 RF Lens(%) = 30 Maximum Injection Time Mode = Auto AGC Target = Standard EASY-IC = Off DataType = Profile Polarity = Positive Source Fragmentation = False Scan Description = Filter IntensityThreshold Intensity Filter Type = IntensityThreshold Minimum Intensity = 5000 Filter DynamicExclusion Share Dynamic exclusion list with other selected dynamic exclusion filters = False Dynamic Exclusion Mode = Custom Exclude after n times = 1 Exclusion duration (s) = 2.5 Excl. Mass Width = ppm Mass tolerance low = 5 Mass tolerance high = 5 Exclude isotopes = True Perform dependent scan on single charge state per precursor only = False Filter TargetedInclusionMassList Add Mass List Targets Determined by Xcalibur AcquireX = False Mass list type = m/z Time Mode = Start/End Time Include Intensity Threshold = True Mass tolerance = ppm Mass tolerance Low = 5 Mass tolerance High = 5 Ignore charge state requirement for unassigned ions = True Set Activation parameters per compound = False Perform dependent scan on most intense ion if no targets are found = True >>>>>>>>>>>>> Mass List Table <<<<<<<<<<<<<< CompoundName| m/z| t start (min)| t stop (min)| Threshold| Caffeine| 195.0877| 0| 5| 0| >>>>>>>>>>>>> End Mass List Table <<<<<<<<<<<<<< Filter TargetedExclusionMassList Filter Label = Xcalibur AcquireX Add Mass List Targets Determined by Xcalibur AcquireX = True Mass list type = m/z Include Intensity Threshold = True Time Mode = Start/End Time Mass tolerance = ppm Mass tolerance Low = 5 Mass tolerance High = 5 Ignore charge state requirement for unassigned ions = True Set Activation parameters per compound = False Perform dependent scan on most intense ion if no targets are found = False >>>>>>>>>>>>> Mass List Table <<<<<<<<<<<<<< CompoundName| m/z| t start (min)| t stop (min)| Threshold| >>>>>>>>>>>>> End Mass List Table <<<<<<<<<<<<<< Data Dependent Properties Data Dependent Mode= Number of Scans Number of Dependent Scans= 5 Scan Event 1 Scan ddMSnScan Multiplex Ions Enabled = False Isolation Offset = Off Isolation Window = Custom Isolation Window (m/z) = 1.5 Maximum number of multiplexed ions = 0 Reported Mass = Offset Mass Collision Energy Type = Normalized HCD Collision Energies (%) = 30,50,70 Orbitrap Resolution = 15000 Time (ms) = 54 Microscans = 1 RF Lens(%) = 50 AGC Target = Standard Maximum Injection Time Mode = Custom EASY-IC = Off DataType = Profile Source Fragmentation = False Scan Description =