Published September 12, 2019 | Version v1
Journal article Open

A Sub-500 μ W Interface Electronics for Bionic Ears

  • 1. METU

Description

This paper presents an ultra-low power current-mode circuit for a bionic ear interface. Piezoelectric (PZT) sensors at the system input transduce sound vibrations into multi-channel electrical signals, which are then processed by the proposed circuit to stimulate the auditory nerves consistently with the input amplitude level. The sensor outputs are first amplified and range-compressed through ultra-low power logarithmic amplifiers (LAs) into AC current waveforms, which are then rectified through custom current-mode circuits. The envelopes of the rectified signals are extracted, and are selectively sampled as reference for the stimulation current generator, armed with a 7-bit user-programmed DAC to enable patient fitting (calibration). Adjusted biphasic stimulation current is delivered to the nerves according to continuous inter-leaved sampling (CIS) stimulation strategy through a switch matrix. Each current pulse is optimized to have an exponentially decaying shape, which leads to reduced supply voltage, and hence ~20% lower stimulator power dissipation. The circuit has been designed and fabricated in 180nm high-voltage CMOS technology with up to 60 dB measured input dynamic range, and up to 1 mA average stimulation current. The 8-channel interface has been validated to be fully functional with 472μW power dissipation, which is the lowest value in the literature to date, when stimulated by a mimicked speech signal.

Files

A_Sub-500__mu_W_Interface_Electronics_for_Bionic_Ears.pdf

Files (2.7 MB)

Additional details

Funding

FLAMENCO – A Fully-Implantable MEMS-Based Autonomous Cochlear Implant 682756
European Commission