Magnetic remanence (spinner) Magnetic remanence was measured on discrete samples by an Agico JR-6A spinner magnetometer, first as natural remanent magnetization (NRM) and then after demagnetization or remagnetization steps were performed on the samples (e.g., alternating field [AF] demagnetization, thermal demagnetization [TD], or isothermal remanent magnetization [IRM]). Data File format Exp: expedition number Site: site number Hole: hole number Core: core number Type: type indicates the coring tool used to recover the core (typical types are F, H, R, X). Sect: section number A/W: archive (A) or working (W) section half. Top offset (cm): top edge of the discrete sample measured relative to the top of a section or section half. Top depth CSF-A (m): location of the observation expressed relative to the top of a hole. Top depth [other] (m): location of the upper edge of the sample expressed relative to the top of the hole. The location is presented in a scale selected by the science party or the report user. Treatment type: type of treatment applied to the discrete specimen. Treatment value (mT or deg C or Agico code): value of the treatment. This may be field or temperature. Inclination (deg): inclination in specimen coordinates computed from JR6 magnetic moment data. Declination (deg): declination in specimen coordinates computed from JR6 magnetic moment data. Inclination geographic (deg): inclination in geographic coordinates computed from JR6 magnetic moment data with correction applied for user entered azimuth and dip. Declination geographic (deg): declination in geographic coordinates computed from JR6 magnetic moment data with correction applied for user entered azimuth and dip. Total intensity (A/m): total intensity as calculated by Rema6 software. Intensity X (A/m): mean intensity value for the x axis as calculated by the Rema6 software. Intensity Y (A/m): mean intensity value for the y axis as calculated by the Rema6 software. Intensity Z (A/m): mean intensity value for the z axis as calculated by the Rema6 software. Azimuth (deg): directional value entered by user at the time of measurement: horizontal angle. Dip (deg): directional value entered by user at time of measurement: vertical angle. Sample type: Indicates the type of sample measured, (e.g., cube). Remasoft TXT file: Remasoft data file. Remasoft JR6ODP file: Remasoft data file with IODP sample names. Remasoft JR6 file: Remasoft data file. This file is necessary for the Rema6 and Remasoft programs to read and write data. Remasoft DAT file: Remasoft data file. Instrument: abbreviation for the instrument used to measure discrete specimen magnetic remanence (JR6). Timestamp (UTC): point in time at which an observation or set of observations was uploaded to the data base. Text ID: automatically generated unique database identifier for a sample, visible on printed labels. Test No: unique number associated with the instrument measurement steps that produced these data. Sample comments: observations recorded about the sample itself. Test comments: observations about a measurement or the measurement process; some measurement observations may be under Result comments. Result comments: observations about a measurement or the measurement process; some measurement observations may be under Test comments.