10.1016/s0304-8853(98)00281-9
https://zenodo.org/records/889513
oai:zenodo.org:889513
Abelmann, Leon
Leon
Abelmann
Porthun, Steffen
Steffen
Porthun
Haast, Marc
Marc
Haast
Lodder, Cock
Cock
Lodder
Moser, Andreas
Andreas
Moser
Best, Margaret E.
Margaret E.
Best
van Schendel, Pieter J. A.
Pieter J. A.
van Schendel
Stiefel, Bruno
Bruno
Stiefel
Hug, Hans J.
Hans J.
Hug
Heydon, Greg P.
Greg P.
Heydon
Farley, Andrew
Andrew
Farley
Hoon, Steve R.
Steve R.
Hoon
Pfaffelhuber, Thomas
Thomas
Pfaffelhuber
Proksch, Roger
Roger
Proksch
Babcock, Ken
Ken
Babcock
Comparing the resolution of magnetic force microscopes using the CAMST reference samples
Zenodo
2015
2015-04-14
Other (Open)
A set of reference samples for comparing the results obtained with different magnetic force microscopes (MFM) has been prepared. These samples consist of CoNi/Pt magneto-optic multilayers with different thicknesses. The magnetic properties of the multilayer are tailored in such a way that a very fine stripe domain structure occurs in remanence. On top of this intrinsic domain structure, bits were written thermomagnetically using different laser powers. These samples have been imaged in six different laboratories employing both home-built and commercial magnetic force microscopes. The resolution obtained with these different microscopes, tips and measurement methods varies between 30 and 100 nm. ; Full-text of this article is not available in this e-prints service. This article was originally published in Journal of magnetism and magnetic materials, published by and copyright Elsevier.