Planned intervention: On Wednesday April 3rd 05:30 UTC Zenodo will be unavailable for up to 2-10 minutes to perform a storage cluster upgrade.
Published April 27, 2009 | Version v1
Technical note Open

Working with weakly magnetized samples on the OPRF cryomagnetometer

  • 1. University of Otago

Description

At the Otago Palaeomagnetic Research Facility we routinely work with very weakly magnetized samples whose remanences are near the limit of the 2G cryomagnetometer’s range. This technical note documents experiments on measurement techniques and parameters and their effects on the precision of the resulting data, and suggests best practices for maximizing measurement precision when using the cryomagnetometer.

Notes

OPRF Booklet 6

Files

working-with-weakly-magnetized-samples.pdf

Files (232.5 kB)

Name Size Download all
md5:da2c933fe6e16cb8b37ff7090c77c4bb
232.5 kB Preview Download