Maringer, Franz Josef
Wiedner, Hannah
Cardellini, Francesco
2019-09-30
<p>In this work, a quick and simple low-level radionuclide metrology technique for the traceable measurement of Rn-222 activity concentration in drinking water using a 0.56 l ionisation chamber operating in spectrometric pulse mode has been developed, tested, verified and applied to 16 water samples successfully. The impact of essential influencing factors on the result has been investigated, discussed and considered in the uncertainty budget of the measurement method. Finally, the new method has been assessed regarding the applicability on the EU Council Directive 2013/51.<em> </em></p>
https://doi.org/10.5281/zenodo.3555047
oai:zenodo.org:3555047
eng
Zenodo
https://doi.org/10.5281/zenodo.3555046
info:eu-repo/semantics/openAccess
Creative Commons Attribution 4.0 International
https://creativecommons.org/licenses/by/4.0/legalcode
Radon 222 activity concentration
traceable calibration
drinking water
radon emanation
ionisation chamber
uncertainty budget
characteristic limits
ISO 11929:2010
ON S 5250-1:2001
AN INNOVATIVE QUICK METHOD FOR TRACABLE MEASUREMENT OF RADON 222 IN DRINKING WATER
info:eu-repo/semantics/article