New TopShim run:

version = topshim 2.0.0
date = 2022-01-25
time = 10:04:24.941
command = topshim 

FINE SHIMMING

Parameters:

probe = Z104275_0392
solvent = CDCl3
shim nucleus = 2H
nucleus optimised for = 1H
o1p (from lock) = 7.24 ppm

Results:


1D Shimming:
Iteration  1: FWHM Before= 2.9 Hz, Target= 0.3 Hz, After= 0.7 Hz
Iteration  2: FWHM Before= 0.7 Hz, Target= 0.3 Hz, After= 0.5 Hz
Iteration  3: FWHM Before= 0.5 Hz, Target= 0.4 Hz, After= 0.7 Hz
Iteration  4: FWHM Before= 0.7 Hz, Target= 0.3 Hz, After= 0.8 Hz
Iteration  5: FWHM Before= 0.8 Hz, Target= 0.4 Hz, After= 1.2 Hz
Iteration  6: FWHM Before= 1.2 Hz, Target= 0.4 Hz, After= 0.6 Hz
Iteration  7: FWHM Before= 0.6 Hz, Target= 0.3 Hz, After= 0.7 Hz
Iteration  8: FWHM Before= 0.7 Hz, Target= 0.3 Hz, After= 0.6 Hz
Slow convergence detected. Shimming was aborted.
Using the best result as the final shim state.

Lineshape quality factor (1D spectrum, spinning) = 0.624

shim changes: 
Z	-430
Z2	-345
Z3	-609
Z4	-3994
Z5	-4745
Z6	-1616

duration = 4 min 48 sec
completed successfully
finished Tue Jan 25 10:09:11 2022


========================================================================