New TopShim run:

version = topshim 2.0.0
date = 2023-01-26
time = 09:15:23.439
command = topshim 

FINE SHIMMING

Parameters:

probe = Z104275_0392
solvent = CDCl3
shim nucleus = 2H
nucleus optimised for = 13C
o1p (from lock) = 7.24 ppm

Results:


1D Shimming:
Iteration  1: FWHM Before= 1.2 Hz, Target= 0.3 Hz, After= 1.9 Hz
Iteration  2: FWHM Before= 1.9 Hz, Target= 0.4 Hz, After= 0.7 Hz
Iteration  3: FWHM Before= 0.7 Hz, Target= 0.3 Hz, After= 0.4 Hz
Iteration  4: FWHM Before= 0.4 Hz, Target= 0.3 Hz, After= 0.5 Hz
Iteration  5: FWHM Before= 0.5 Hz, Target= 0.3 Hz, After= 0.4 Hz
Iteration  6: FWHM Before= 0.4 Hz, Target= 0.3 Hz, After= 0.5 Hz
Iteration  7: FWHM Before= 0.5 Hz, Target= 0.4 Hz, After= 0.4 Hz
Slow convergence detected. Shimming was aborted.
Using the best result as the final shim state.

Lineshape quality factor (1D spectrum, spinning) = 0.507

shim changes: 
Z	-246
Z2	-389
Z3	-427
Z4	-4818
Z5	-7694
Z6	-4773

duration = 3 min 32 sec
completed successfully
finished Thu Jan 26 09:18:55 2023


========================================================================