New TopShim run:

version = topshim 1.8.0
date = 2020-09-29
time = 13:19:19.030
command = topshim  z6off 1d optsolvent coil=1H tuneboff tunea

1D SHIMMING

Parameters:

maximum order = 5
probe = Z161014_0003
solvent = CDCl3
shim nucleus = 2H
nucleus optimised for = 1H
o1p (from lock) = 7.24 ppm
optimisation parameters = ls
linewidth 1H = 0.10 Hz
envelope shape / strictness = 2.00 / 200

Results:

echo time must be reduced
initial B0 stdDev = 13.43 Hz
sample size = 2.23 cm, position = -0.10 cm
final B0 stdDev = 3.31 Hz > improvement = 4.1
envelope width = 1.04 Hz
tune lock level increase = 1.0
shim changes: 
Z	+100
Z2	+38
Z3	-168
Z4	-3068
Z5	-3413
YZ	+55

duration = 6 min 9 sec
completed successfully
finished Tue Sep 29 13:25:25 2020


========================================================================