# 1 "C:/Bruker/TopSpin3.5pl6/exp/stan/nmr/lists/pp/cp90"
;cp90
;
;TS3 / 03.06.2011
;
;basic cp experiment with flipback 90 at pl11
;written by HF 1.3.2001
;
;checked by SEWE 02.08.2011
;
;
;Avance III version
;parameters: 
;pl1 : f1 power level for CP
;pl2 : =0W, not used
;pl11 : to determine X 90 or 180 degree pulse
;pl12 : f2 power level for 90 deg + decoupling 
;pl13 : f2 power level in case of special decoupling (e.g. spinal64_13)
;sp0 : proton power level during contact
;p1 : f1 90 or 180 deg pulse at plw11
;p3 : proton 90 at power level plw12
;p15 : contact time at plw1 (f1) and spw0 (f2)
;cpdprg2 : cw, spinal64 (at plw12)
;d1 : recycle delay
;pcpd2 : pulse length in decoupling sequence (e.g. 180-0.2us)
;spnam0 : use e.g. ramp.100 for variable amplitude CP
;zgoptns : -Dfslg, -Dlacq, -Dlcp15, or blank
;
;
;$CLASS=Solids
;$DIM=1D
;$TYPE=cross polarisation
;$SUBTYPE=simple 1D
;$COMMENT=basic cp experiment, flipback pulse on X for 90-degree pulse calibration

prosol relations=<solids_cp>

# 1 "C:/Bruker/TopSpin3.5pl6/exp/stan/nmr/lists/pp/Avancesolids.incl" 1
;Avancesolids.incl
;   for TCU3
;
;version 1 28.2.2001, written by HF
;	   4.4.2006, complemented by Ste
# 7 "C:/Bruker/TopSpin3.5pl6/exp/stan/nmr/lists/pp/Avancesolids.incl"
;$CLASS=Solids INCL
;$COMMENT=definitions for setrtp and setnmr

;switch between 1H or 19F output (H amplifier) 

# 14 "C:/Bruker/TopSpin3.5pl6/exp/stan/nmr/lists/pp/Avancesolids.incl"
;gating pulse for RX, ADC and HPPR auto mode


# 18 "C:/Bruker/TopSpin3.5pl6/exp/stan/nmr/lists/pp/Avancesolids.incl"
;explicit blanking of preamp, receiver, adc, not implemented yet


# 26 "C:/Bruker/TopSpin3.5pl6/exp/stan/nmr/lists/pp/Avancesolids.incl"
;generate dwell clock


;turn dwell clock on/off
# 32 "C:/Bruker/TopSpin3.5pl6/exp/stan/nmr/lists/pp/Avancesolids.incl"
;


# 36 "C:/Bruker/TopSpin3.5pl6/exp/stan/nmr/lists/pp/Avancesolids.incl"
;define data sampling for WAHUHA type experiments, samples analogue or digitally for d9, closes receiver


# 40 "C:/Bruker/TopSpin3.5pl6/exp/stan/nmr/lists/pp/Avancesolids.incl"
;explicit transmitter blanking, not supported yet

;NCO switching, to be implemented

;$id: $
# 37 "C:/Bruker/TopSpin3.5pl6/exp/stan/nmr/lists/pp/cp90" 2
# 38 "C:/Bruker/TopSpin3.5pl6/exp/stan/nmr/lists/pp/cp90"
"acqt0=0" ; value for baseopt calculation

# 1 "mc_line 40 file C:/Bruker/TopSpin3.5pl6/exp/stan/nmr/lists/pp/cp90 dc-measurement inserted automatically"

    dccorr
# 40 "C:/Bruker/TopSpin3.5pl6/exp/stan/nmr/lists/pp/cp90"
1 ze
2 d1 do:f2 

# 1 "C:/Bruker/TopSpin3.5pl6/exp/stan/nmr/lists/pp/p15_prot.incl" 1
;p15_prot.incl
;avance-version (13/01/29)
;Test the rf on-time and duty cycle
;protect against too long contact time
;Test the rf on-time and duty cycle 
;
;$CLASS=Solids INCL
;$COMMENT=protect against too long contact time
# 11 "C:/Bruker/TopSpin3.5pl6/exp/stan/nmr/lists/pp/p15_prot.incl"
1m
if "p15 < 10.1m" goto Passp15
2u
print "contact time exceeds 10msec limit!"
goto HaltAcqu
Passp15, 1m


# 20 "C:/Bruker/TopSpin3.5pl6/exp/stan/nmr/lists/pp/p15_prot.incl"
;$Id: p15_prot.incl,v 1.8 2013/01/29 15:40:14 ber Exp $
# 43 "C:/Bruker/TopSpin3.5pl6/exp/stan/nmr/lists/pp/cp90" 2

			;make sure p15 does not exceed 10 msec 
			;let supervisor change this pulseprogram if 
			;more is needed

# 50 "C:/Bruker/TopSpin3.5pl6/exp/stan/nmr/lists/pp/cp90"
			;disable protection file for long acquisition change decoupling power !!! or you risk probe damage
			;if you set the label lacq (ZGOPTNS -Dlacq), the protection is disabled

# 1 "C:/Bruker/TopSpin3.5pl6/exp/stan/nmr/lists/pp/aq_prot.incl" 1
;aq_prot.incl
;avance-version (13/01/28)
;Test the rf on-time and duty cycle
;protect against too long acquisition time

;$CLASS=Solids INCL
;$COMMENT=protect against too long acquisition time
# 9 "C:/Bruker/TopSpin3.5pl6/exp/stan/nmr/lists/pp/aq_prot.incl"
1m
if "aq < 50.1m" goto Passaq
2u 
print "acquisition time exceeds 50m limit!"
goto HaltAcqu
Passaq, 1m

;$Id: aq_prot.incl,v 1.8 2013/01/28 17:39:02 ber Exp $
# 53 "C:/Bruker/TopSpin3.5pl6/exp/stan/nmr/lists/pp/cp90" 2

			;allows max. 50 msec acquisition time, supervisor
			;may change  to max. 1s at less than 5 % duty cycle
			;and reduced decoupling field
# 59 "C:/Bruker/TopSpin3.5pl6/exp/stan/nmr/lists/pp/cp90"
  1u fq=cnst21:f2
  (p3 pl12 ph1):f2
  (p15 pl1 ph2):f1 (p15:sp0 ph10):f2
  (p1 ph3):f1 (1u pl11):f1 (1u cpds2):f2
  go=2 ph31
  1m do:f2
  wr #0
HaltAcqu, 1m
exit

ph0= 0
ph1= 1 3
ph2= 0 0 2 2 1 1 3 3
ph3= 1 1 3 3 2 2 0 0
ph10= 0
ph31= 0 2 2 0 1 3 3 1

# 78 "C:/Bruker/TopSpin3.5pl6/exp/stan/nmr/lists/pp/cp90"
;$Id: cp90,v 1.7 2012/10/09 09:17:28 ber Exp $